Automatic Flash point closed cup tester Tag Lintel ATVT-20

full automation of the testing process of flash point

SKU: ATVT-20 Category: Tag:
Description
  • The Lintel ATVT-20 device provides full automation of the testing process – filling the bath with water, maintaining the heating rate of the product, testing the product with built-in electric spark ignition, determining and storing the flash point with automatic adjustment to atmospheric pressure, replacing a heated coolant with a cold one
  • Automatic coolant renewal after the end of the test reduces the time between tests
  • Integration with the LinteL Link data collection system allows you to collect and transfer laboratory test results from the device to a personal computer via wireless communication. Integration with the laboratory information system Lintel LIS provides comprehensive automation of laboratory activities
  • High-contrast color TFT display provides an intuitive user interface with all the information you need while testing and viewing results
  • Test log maintenance provides storage in non-volatile memory of up to 100 test results (test method, product name, expected flash point, flash point, atmospheric pressure, heating rate, date and time of test completion, performer), as well as a graph of heating rate during the last test
  • Automatic stop of the test with a sound signal about the end of the test
  • Detection of product overheating with automatic end of test
  • High precision product temperature sensor
  • Full self-diagnosis system with indication of the causes of malfunctions on the display

EQUIPMENT:

  1. LinteL ATVT-20 device – 1 pc.;
  2. Crucible cover – 1 pc.;
  3. Crucible – 2 pcs.;
  4. Screw clamp 12-20 – 4 pcs.;
  5. Reinforced tube 1/2″ 1 m – 2 pcs.;
  6. Pressure reduction valve Honeywell D04FS-1/2″ A – 1 pc.;
  7. Operation manual – 1 pc.;
  8. Passport – 1 pc.
Additional information
Flash point detection rangefrom +13 to +93 °C
Product temperature measurement resolution0.1 °C
Product heating rate, depending on the selected test conditions1±0.1°C/min, 3±0.3 °C/min
Atmosphere pressurefrom 84 to 106.6 kPa (from 630 to 800 mm Hg)
Refrigerant temperature (water)from + 4 to +15 °C
Refrigerant flow (water)from 2.5 to 5.0 l/min
Heating rate for products with a flash point below plus 60°Cfrom 0.9 to 1.1 °C/min
Heating speed for products with a flash point of plus 60°C and abovefrom 2.7 to 3.3 °C/min
Atmospheric pressure sensor errormmHg. ±(12.0+α)*
Power consumptionno more than 500 VA
Device dimensions370 x 420 x 310 mm (depth x width x height)
Device weightno more than 14 kg
Ambient temperaturefrom +15 to +35 °C
Relative humidityno more than 75%
Mains voltagefrom 187 to 242 V
Power frequency49 to 51 Hz
Life time6 years, no more than 15,000 hours
Warranty period1 year, no more than 2,500 hours
StandardsASTM D 56, GOST R 53717-2009